Página 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL0345 200PYRGEOMETERmanualCustomer Support

Página 2 - IMPORTANT USER INFORMATION

GENERAL INFORMATION 9 Within this spectral range the earth is able to maintain an equilibrium temperature by losing a certain quantity of heat

Página 3 - DECLARATION OF CONFORMITY

GENERAL INFORMATION 10 Figure 1.3 Atmospheric radiation (atmospheric window 8 to 14 µm) Where as a pyranometer only receives

Página 4

GENERAL INFORMATION 11 = Upward irradiance of the CG 4 sensor [W/m2] Note that the net radiation term (Uemf / S) is mostly negative, so th

Página 5

GENERAL INFORMATION 12 • The thermistor contacts a part of the window, it is a blackbody radiator and heat source itself and its materi

Página 6

GENERAL INFORMATION 13 1.4.1 How to perform the check? The check must be performed under clear sky conditions. The CG 4 pyrgeometer is opera

Página 7

GENERAL INFORMATION 14CG 4 stabilises at < 3 W/m2. The experiment shows that for most practical purposes, CG 4 does not need compensation fo

Página 8

GENERAL INFORMATION 15 Due to the thermopile construction and applied thermistor compensation circuit the temperature response is suppressed to

Página 9

GENERAL INFORMATION 16

Página 10

TECHNICAL DATA 17 2 TECHNICAL DATA 2.1 SPECIFICATIONS OF THE CG 4 PYRGEOMETER Performance Spectral range: 4.5 to 42 µm, 50% points. Se

Página 11 - 1067.5 ⋅⋅+=

TECHNICAL DATA 18 Estimated inaccuracy of measurement: < 7.5 W/m2. Thermistor specifications (only for thermistor version): Type YSI 44

Página 12 - 1067.5 ⋅⋅

IMPORTANT USER INFORMATION 1 IMPORTANT USER INFORMATION Reading this entire manual is recommended for full understanding of the use of this

Página 13

TECHNICAL DATA 19 Figure 2.1 CG 4 pyrgeometer outline dimensions

Página 14

TECHNICAL DATA 20 2.2 ACCURACY As listed in paragraph 2.1 the sensitivity is cross-correlated to a number of parameters such as temperature a

Página 15

TECHNICAL DATA 21 The temperature dependence of the sensitivity is an individual function. For any given CG 4 the curve lies in the region betw

Página 16

TECHNICAL DATA 22

Página 17

INSTALLATION 23 3 INSTALLATION Reading these instructions before installation is recommended. 3.1 DELIVERY Check the contents of the shipm

Página 18 - Max. 4 W/m

INSTALLATION 24 3.2 MECHANICAL INSTALLATION The mechanical installation of the pyrgeometer must be carried out depending on the application. D

Página 19

INSTALLATION 25 Note: After recalibration and/or reinstallation the nylon insulators must be replaced with new ones to maintain durability.

Página 20

INSTALLATION 26 3.2.3 Levelling Accurate measurement of the downward atmospheric long wave radiation requires the proper levelling of the ther

Página 21 - Non- linearity

INSTALLATION 27 Figure 3.2 Mounting of two pyrgeometers. The pyrgeometer must always be mounted with the cable gla

Página 22 - (Typical curves)

INSTALLATION 28 3.3 ELECTRICAL CONNECTION The CG 4 is provided with a 10 m, 6 wire shielded cable (8 wire for the Pt-100 version). The followi

Página 23

IMPORTANT USER INFORMATION 2 DECLARATION OF CONFORMITY According to EC guideline 89/336/EEC 73/23/EEC We Kipp & Zonen B.V. Röntgen

Página 24

INSTALLATION 29 Looking at the circuit diagram of figure 3.3, it is clear that the impedance of the readout equipment is loading the thermistor

Página 25

INSTALLATION 30Kipp & Zonen supplies shielded low-noise extension cables up to lengths of 200 m which are coupled by waterproof connectors

Página 26

INSTALLATION 31 Figure 3.5 Example of a CG 4 with a Pt-100 connected to a Campbell data logger For amplification of the p

Página 28

OPERATION 33 4 OPERATION After completing the installation the pyrgeometer will be ready for operation. At the end of the paragraph an exampl

Página 29

OPERATION 34 4.1.1 Example During field measurements the pyrgeometer is exposed to varying atmospheric conditions with typical radiating prop

Página 30 - Ω or more such as

OPERATION 35 4.1.3 Clear sky conditions Clear sky conditions differ in the way that there is a relative large heat loss caused by the atmosph

Página 31

OPERATION 36on many parameters, for example the water vapour content in the atmosphere (Humidity), location of the CG 4 at a certain altitude a

Página 32

OPERATION 37 4.2 MEASURING NET RADIATION WITH TWO PYRGEOMETERS With two CG 4’s as a net-pyrgeometer, the long wave radiation balance (also

Página 33

OPERATION 38

Página 34

TABLE OF CONTENTS 3 TABLE OF CONTENTS IMPORTANT USER INFORMATION ...1 DECLARATION OF CONF

Página 35

MAINTENANCE 39 5 MAINTENANCE Once installed the pyrgeometer needs little maintenance. The Silicon window must be inspected at regular inte

Página 36

MAINTENANCE 40 It is normal in humid areas to replace the desiccant twice a year. The exchange interval is affected by humidity, change in ai

Página 37 - mid of September

CALIBRATION 41 6 CALIBRATION 6.1 THE CALIBRATION FACTOR The ideal pyrgeometer should always have a constant ratio of voltage output to irra

Página 38

CALIBRATION 42 A final St is determined only from one minute St’s determined in periods with net IR-exchange = Ld – Tb481067.5 ⋅⋅−> 40 W/m²

Página 39

FREQUENTLY ASKED QUESTIONS (FAQ’S) 43 7 FREQUENTLY ASKED QUESTIONS (FAQ’s) The most frequently asked questions are listed below. For an

Página 40 - MAINTENANCE

FREQUENTLY ASKED QUESTIONS (FAQ’S) 44 normally the cable gland is never touched while the cartridge is removed frequently. So when no care is

Página 41

TROUBLE SHOOTING 45 8 TROUBLE SHOOTING Any visible damage or malfunction should be reported to your dealer, who will suggest appropriate acti

Página 42

TROUBLE SHOOTING 46

Página 43

PART NUMBERS / SPARE PARTS / OPTIONS 47 9 PART NUMBERS / SPARE PARTS / OPTIONS Description Part no. Upper sunscreen (plastic) 0305-166

Página 44 - (Lnet = -150 W/m

PART NUMBERS / SPARE PARTS / OPTIONS 48Description Part no. Mounting plate for 4 unventilated sensors (2 upper and 2 lower) 0012 092 10

Página 45

TABLE OF CONTENTS 4 6.2 CALIBRATION PROCEDURE AT KIPP & ZONEN ... 41 6.2.1 The outdoor procedure...

Página 46

APPENDIX I 49 APPENDIX I WORLD RADIATION CENTRE INFORMATION The World Radiation Centre capable of Pyrgeometer calibration is: Physikalis

Página 47

APPENDIX I 50

Página 48

APPENDIX II 51 APPENDIX II THERMISTOR SPECIFICATIONS YSI thermistor 44031 Resistance versus Temperature in °C Temperature [ °C ] Resistan

Página 49

APPENDIX II 52

Página 50

APPENDIX III 53 APPENDIX III Pt-100 SPECIFICATIONS Pt-100 Resistance versus Temperature in °C Temperature [ °C ] Resistance [ Ω ] Temperat

Página 51

APPENDIX III 54

Página 52

APPENDIX VI 55 APPENDIX IV RECALIBRATION SERVICE Pyranometers, UV-meters, Pyrgeometers & Sunshine duration sensors Kipp & Zonen s

Página 53

APPENDIX VI 56

Página 54

APPENDIX VI 57 NAME : COMPANY/INSTITUTE : ADDRESS : POSTCODE +CITY : COUNTRY : PHONE : FAX :  I would like to receive a price list for r

Página 55

APPENDIX VI 58

Página 56

GENERAL INFORMATION 5 1 GENERAL INFORMATION 1.1 INTRODUCTION Pyrgeometers are recommended by the BSRN (Baseline Surface Radiation Network)

Página 57

APPENDIX VI 59 NAME : COMPANY/INSTITUTE : ADDRESS : POSTCODE +CITY : COUNTRY : PHONE : FAX :  I would like to receive a price list for rec

Página 59

APPENDIX VI 61 NAME : COMPANY/INSTITUTE : ADDRESS : POSTCODE +CITY : COUNTRY : PHONE : FAX :  I would like to receive a price list for r

Página 60

APPENDIX VI 62

Página 61

Holland Kipp & Zonen B.V. Röntgenweg 1 2624 BD DELFT T +31 15 269 8000 F +31 15 262 0351 E [email protected] Kipp & Zonen

Página 62

GENERAL INFORMATION 6 1.2 PHYSICAL PRINCIPLES OF THE PYRGEOMETER The CG 4 pyrgeometer is provided with a thermal detector. The thermal detec

Página 63

GENERAL INFORMATION 7 1.2.1 Properties of the Silicon window CG 4 uses a specially designed pure silicon window. Although the window isn’t h

Página 64

GENERAL INFORMATION 8 Figure 1.2 Typical transmittance of a CG 4 window. 1.3 DOWNWARD ATMOSPHERIC LONG WAVE RADIATION Th